Work

Transcranial photoacoustic tomography of the monkey brain

Year: 2012
Type: conference-paper
Abstract: A photoacoustic tomography (PAT) system using a virtual point ultrasonic transducer was developed for transcranial imaging of monkey brains. The virtual point transducer provided a 10 times greater field-of-view (FOV) than finiteaperture unfocused transducers, which enables large primate imaging. The cerebral cortex of a monkey brain was accurately mapped transcranially, through up to two skulls ranging from 4 to 8 mm in thickness. The mass density and speed of sound distributions of the skull were estimated from adjunct X-ray CT image data and utilized with a timereversal algorithm to mitigate artifacts in the reconstructed image due to acoustic aberration. The oxygenation saturation (sO2) in blood phantoms through a monkey skull was also imaged and quantified, with results consistent with measurements by a gas analyzer. The oxygenation saturation (sO2) in blood phantoms through a monkey skull was also imaged and quantified, with results consistent with measurements by a gas analyzer. Our experimental results demonstrate that PAT can overcome the optical and ultrasound attenuation of a relatively thick skull, and the imaging aberration caused by skull can be corrected to a great extent.
Language: English
Open Access status: bronze
OpenAlex ID: W2028953622
FWCI: 2.061
Cites: 20
Cited by: 3
Related to: 10