Work

A determination of Planck's constant using the NRC watt balance

Year: 2014
Type: article
Abstract: We have measured Planck's constant and have obtained a value of 6.626 070 34(12) × 10 −34 J s. To our knowledge this measurement of h has the lowest uncertainty reported to date. This result has been obtained from measurements of four masses of different material and nominal values varying from 1 kg to 250 g. The experimental procedures and the measurement uncertainties are described in detail.
Source: Metrologia
Language: English
Open Access status: bronze
OpenAlex ID: W2033787185
FWCI: 21.2
Cites: 23
Cited by: 93
Related to: 10