Work

Is TrueFISP a gradient‐echo or a spin‐echo sequence?

Year: 2003
Type: article
Abstract: It is commonly accepted that TrueFISP (balanced FFE, FIESTA) belongs to the class of gradient-echo (GRE) sequences. GRE sequences are sensitive to dephasing effects of the transverse magnetization between the excitation pulse and echo acquisition, and phase coherence is only established directly after and before excitation pulses. However, an analysis of the phase evolution of transverse magnetization in a TrueFISP experiment shows very close similarities to the echo formation of a spin-echo (SE) experiment. If dephasing between excitation pulses is below +/-pi, TrueFISP exhibits a nearly complete refocusing of transverse magnetization at TE = TR/2. Only signals acquired before and after TR/2 show an additional T*2 sensitivity.
Language: English
Field: Medicine
Open Access status: bronze
OpenAlex ID: W2035519549
FWCI: 10.74
Cites: 9
Cited by: 202
Related to: 10