Year: 1974
Type: article
Abstract: Modifications are described to the Hewlett‐Packard model 185B CHN analyzer. “Particulate” organic carbon and nitrogen on 24‐mm‐diam GFC filters can easily be run simultaneously with the modified analyzer. With appropriate blank controls, the practical lower limits for analysis are about 5 µg C and 1 µg N. Thus it is possible to filter 1 liter or less for all seawater samples other than deep oceanic ones where up to 5 liters would probably be preferable.
Source: Limnology and Oceanography
Author: Jonathan H. Sharp
Institution: University of California San Diego
Language: English
Subfield: Oceanography
Field: Earth and Planetary Sciences
Domain: Physical Sciences
Open Access status: bronze