Application of single-image camera calibration for ultrasound augmented laparoscopic visualization
Year: 2015
Type: conference-paper
Abstract: laparoscopes obtained using rdCalib and the well-accepted OpenCV camera calibration method were compared. Target registration error (TRE) as a measure of camera calibration accuracy for our optical tracking-based AR system was also compared between the two calibration methods. Based on our experiments, the single-image camera calibration yields consistent and accurate results (mean TRE = 1.18 ± 0.35 mm for the 5-mm scope and mean TRE = 1.13 ± 0.32 mm for the 10-mm scope), which are comparable to the results obtained using the OpenCV method with 30 images. The new single-image camera calibration method is promising to be applied to our augmented reality visualization system for laparoscopic surgery.
DOI: 10.1117/12.2082194
Language: English
Subfield: Computer Vision and Pattern Recognition
Field: Computer Science
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W2035914531