Work

Monte Carlo dose calculations for dynamic IMRT treatments

Year: 2001
Type: article
Abstract: Dose calculations for intensity modulated radiation therapy (IMRT) face new challenges due to the complex leaf geometry and time dependent nature of the delivery. A fast method of particle transport through a dynamic multileaf collimator (MLC) geometry that accounts for photon attenuation and first-scattered Compton photon production has been incorporated into an existing Monte Carlo code used for patient dose calculations. Dosimetric agreement between calculation and measurement for two photon energies and MLC types is within experimental error for the sliding window tests. For a patient IMRT field, the Monte Carlo calculations are closer to measured dose than similar superposition or pencil beam calculations.
Language: English
Subfield: Radiation
Open Access status: bronze
OpenAlex ID: W2036945320
FWCI: 7.473
Cites: 19
Cited by: 81
Related to: 10