Work

Mitigation of susceptibility‐induced signal loss in neuroimaging using localized shim coils

Year: 2005
Type: article
Abstract: Correction of magnetic field distortions is essential for obtaining accurate brain blood-oxygen-level-dependent functional magnetic resonance imaging (fMRI) activation maps. The present work introduces an active shimming method that utilizes the magnetic field generated by resistive shim coils placed in the mouth to locally homogenize the magnetic field in the inferior portion of the frontal lobe, where the field is most seriously distorted. The shimming field can be optimized in situ patient by patient for the region of interest of the scanner operator's choice. The method at 1.5 T is shown to be effective in reducing field inhomogeneity and in recovery of fMRI signal. For example, in a region of interest approximately of 149 cm3, a coil of simple geometry can reduce the root mean square of the magnetic field by more than 50% and the recovered signal increases the extent of activation detected in a breath-holding fMRI experiment.
Language: English
Field: Medicine
Open Access status: bronze
OpenAlex ID: W2043674983
FWCI: 3.639
Cites: 12
Cited by: 102
Related to: 10