Work

Overtone Mobility Spectrometry: Part 4. OMS-OMS Analyses of Complex Mixtures

Year: 2011
Type: article
Abstract: A new, two-dimensional overtone mobility spectrometry (OMS-OMS) instrument is described for the analysis of complex peptide mixtures. OMS separations are based on the differences in mobilities of ions in the gas phase. The method utilizes multiple drift regions with modulated drift fields such that only ions with appropriate mobilities are transmitted to the detector. Here we describe a hybrid OMS-OMS combination that utilizes two independently operated OMS regions that are separated by an ion activation region. Mobility-selected ions from the first OMS region are exposed to energizing collisions and may undergo structural transitions before entering the second OMS region. This method generates additional peak capacity and allows for higher selectivity compared with the one-dimensional OMS method. We demonstrate the approach using a three-protein tryptic digest spiked with the peptide Substance P. The [M + 3H](3+) ion from Substance P can be completely isolated from other components in this complex mixture prior to introduction into the mass spectrometer.
Language: English
Subfield: Spectroscopy
Field: Chemistry
Open Access status: green
OpenAlex ID: W2048781201
FWCI: 1.304
Cites: 50
Cited by: 20
Related to: 10