Year: 2005
Type: article
Abstract: The human lung and its functions are extremely sensitive to gravity; however, the conventional high-field magnets used for most laser-polarized (3)He MRI of the human lung restrict subjects to lying horizontally. Imaging of human lungs using inhaled laser-polarized (3)He gas is demonstrated in an open-access very-low-magnetic-field (<5 mT) MRI instrument. This prototype device employs a simple, low-cost electromagnet, with an open geometry that allows variation of the orientation of the imaging subject in a two-dimensional plane. As a demonstration, two-dimensional lung images were acquired with 4-mm in-plane resolution from a subject in two orientations: lying supine and sitting in a vertical position with one arm raised. Experience with this prototype device will guide optimization of a second-generation very-low-field imager to enable studies of human pulmonary physiology as a function of subject orientation.
Source: Magnetic Resonance in Medicine
DOI: 10.1002/mrm.20456
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $3,680