Work

Effects of respiratory motion on dose uniformity with a charged particle scanning method

Year: 1992
Type: article
Abstract: A three-dimensional spot-scanning technique for radiotherapy with protons is being developed at the Paul Scherrer Institute. As part of the effort to optimize the design and ensure clinically useful dose distributions, a computer simulation of the dose deposition in the presence of respiratory motion was performed. Preliminary experiments have characterized the proton beam and the scanning procedure. Using these parameters, the computer program calculated the dose within a uniform volume of water in the presence of respiratory motion. Respiration amplitude, respiration period, respiration direction, number of fractions, size and position of the beamspots and rescanning multiplicity were systematically varied and the effect on the dose distribution determined. The dose uniformity is very dependent on the direction of the respiration relative to the three independent beam scanning directions. The dose uniformity decreases with increasing respiration amplitude, but has little response to changes in respiration frequency. Rescanning the volume, such as with fractionation, improves the dose uniformity roughly as the square root of the number of fractions. Broad, Gaussian beams result in better dose uniformity than narrow, sharply delineated ones, but produce slower dose fall-off at the edges of the scanned volume. Results of this work are being incorporated into the design of the system.
Language: English
Field: Medicine
Open Access status: bronze
OpenAlex ID: W2059950254
FWCI: 1.403
Cites: 14
Cited by: 254
Related to: 10