Year: 2010
Type: conference-paper
Abstract: In this study, we further developed our photoacoustic endoscopic system to produce three-dimensional images of internal organs by performing pullback C-scans. Employing the side-scanning photoacoustic endoscopic probe discussed in the Optical Society of America's journal Optics Letters, we could acquire successive B-scan images by pulling back the probe with a motorized linear stage. We demonstrate the endoscopic system's volumetric imaging ability through imaging of a metal wire phantom and an in situ rat rectum.
DOI: 10.1117/12.842503
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: bronze