Work

High-resolution coregistered intravascular imaging with integrated ultrasound and optical coherence tomography probe

Year: 2010
Type: article
Abstract: We report an integrated ultrasound (US) and optical coherence tomography (OCT) probe and system for intravascular imaging. The dual-function probe is based on a 50 MHz focused ring US transducer, with a centric hole for mounting OCT probe. The coaxial US and light beams are steered by a 45° mirror to enable coregistered US∕OCT imaging simultaneously. Lateral resolution of US is improved due to focused ultrasonic beam. Mirror effects on US were investigated and invitro imaging of a rabbit aorta has been carried out. The combined US-OCT system demonstrated high resolution in visualizing superficial arterial structures while retaining deep penetration of ultrasonic imaging.
Language: English
Open Access status: bronze
OpenAlex ID: W2063052681
FWCI: 4.908
Cites: 12
Cited by: 83
Related to: 10