Year: 2007
Type: article
Abstract: An imaging method called "quantification of relaxation times and proton density by twin-echo saturation-recovery turbo-field echo" (QRAPTEST) is presented as a means of quickly determining the longitudinal T(1) and transverse T(2) (*) relaxation time and proton density (PD) within a single sequence. The method also includes an estimation of the B(1) field inhomogeneity. High-resolution images covering large volumes can be achieved within clinically acceptable times of 5-10 min. The range of accuracy for determining T(1), T(2) (*), and PD values is flexible and can be optimized relative to any anticipated values. We validated the experimental results against existing methods, and provide a clinical example in which quantification of the whole brain using 1.5 mm(3) voxels was achieved in less than 8 min.
Source: Magnetic Resonance in Medicine
DOI: 10.1002/mrm.21165
Authors: J.B.M. Warntjes, O. Dahlqvist, P. Lundberg
Institutions: Linköping University, Linköping University Hospital
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W2068155703