Year: 2013
Type: conference-paper
Abstract: The Grüneisen parameter of tissue is a constitutive parameter in photoacoustic tomography. Here, we applied photoacoustic spectrometry (PAS) to directly measure the Grüneisen parameter. In our PAS system, laser pulses at wavelengths between 460 and 1600 nm were delivered to tissue samples, and photoacoustic signals were detected by a 20 MHz flat water-immersion ultrasonic transducer. By fitting photoacoustic spectra to light absorption spectra, we found that the Grüneisen parameter was 0.73 for porcine subcutaneous fat tissue, and 0.15 for oxygenated bovine red blood cells at room temperature (24°C).
DOI: 10.1117/12.2004117
Authors: Da-Kang Yao, Lihong V. Wang
Institution: Washington University in St. Louis
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: bronze
OpenAlex ID: W2071248337