Year: 2002
Type: article
Abstract: Three-dimensional T(1)-weighted magnetization-prepared rapid gradient-echo (MP-RAGE) sequences with centric phase encoding (PE) in the inner loop provide structural brain images with a high spatial resolution and high tissue contrast. A disadvantage of this sequence type is the susceptibility to inhomogeneities of the radiofrequency (RF) coil, which may result in poor image contrast in some peripheral regions. A special excitation pulse is presented which compensates for these effects in both the head/foot and anterior/posterior directions. This pulse has a duration of only 1.3 ms and is thus compatible with the short repetition times (TRs) required for MP-RAGE imaging. It is shown experimentally that images acquired with the compensation pulse may be segmented without using intensity correction algorithms during data postprocessing.
Source: Magnetic Resonance in Medicine
DOI: 10.1002/mrm.10050
Authors: R. Deichmann, C.D. Good, R. Turner
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W2071617502
