Work

Experimental validation of theoretical models for the frequency response of atomic force microscope cantilever beams immersed in fluids

Year: 2000
Type: article
Abstract: Detailed measurements of the frequency responses of a series of rectangular atomic force microscope (AFM) cantilever beams, immersed in a range of fluids, have been performed to test the validity and accuracy of the recent theoretical model of Sader [J. Appl. Phys. 84, 64 (1998)]. This theoretical model gives the frequency response of a cantilever beam, that is immersed in a viscous fluid and excited by an arbitrary driving force. Very good agreement between experimental measurements and theoretical calculations is found for all fluids considered. Furthermore, a critical assessment of the well-known inviscid model is presented, which demonstrates that this model is not applicable to AFM cantilever beams in general.
Language: English
Open Access status: bronze
OpenAlex ID: W2084403547
FWCI: 3.747
Cites: 21
Cited by: 335
Related to: 10