Fatigue properties of cast and heat treated Ti6Al4V alloy for anatomic hip prostheses
Year: 1987
Type: article
Abstract: The intricate shape of Ti-6Al-4V anatomic hip prostheses necessitates the use of casting as a fabrication method, since neither machining, nor forging or powder metallurgial processing are reasonably acceptable. Cast and hot isostatically pressed Ti-6Al-4V, however, has a relatively low ductility and reduced fatigue properties. Appropriate thermal treatments can improve these properties. Specifically, treatments comprising a beta-solutionizing and an aging step yield ductilities similar to the forged and annealed condition, and fatigue properties 11.5% above the non-treated condition. Such treatments produce small alpha-beta platelet colonies with a ragged, tortuous morphology. The reduction of scatter in fatigue strength is the result of the formation of this well controlled microstructure.
Source: Biomaterials
Authors: P. Ducheyne, D. Kohn, T.S. Smith
Institution: University of Pennsylvania
Language: English
Subfield: Materials Chemistry
Field: Materials Science
Domain: Physical Sciences
Open Access status: hybrid
Estimated APC paid: $4,640
OpenAlex ID: W2085756274
