Fiber-optic catheter-based polarization-sensitive OCT for radio-frequency ablation monitoring
Year: 2014
Type: article
Abstract: An all-fiber optic catheter-based polarization-sensitive optical coherence tomography system is demonstrated. A novel multiplexing method was used to illuminate the sample, splitting the light from a 58.5 kHz Fourier-domain mode-locked laser such that two different polarization states, alternated in time, are generated by two semiconductor optical amplifiers. A 2.3 mm forward-view cone-scanning catheter probe was designed, fabricated, and used to acquire sample scattering intensity and phase retardation images. The system was first verified with a quarter-wave plate and then by obtaining intensity and phase retardation images of high-birefringence plastic, human skin in vivo, and untreated and thermally ablated porcine myocardium ex vivo. The system can potentially in vivo image of the cardiac wall to aid radio-frequency ablation therapy for cardiac arrhythmias.
Source: Optics Letters
DOI: 10.1364/ol.39.005066
Institution: Case Western Reserve University
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green