Point-of-Care Measurements of HbA1c: Simplicity Does Not Mean Laxity With Controls

Year: 2012
Type: article
Study design: Observational Study
Abstract: Point-of-care HbA1c measurements(POC-A1Cs) have been adopted bymany diabetes clinics to improve the quality of care provided to their pa-tients (1). Herein, we show that reliability of this approach might be questioned. POC-A1Cs routinely used in the ambula-tory section of our diabetes clinic was evaluated on 100 diabetic patients (type 1, n 5 58; type 2, n 5 42) attending the clinic from 1 October 2011 to 30 Novem-ber 2011. Patients with abnormal hemo-globin traits or shortened erythrocyte life span were excluded. Blood-capillary samples were analyzed by POC-A1C (DCA Vantage; Siemens Medical Solu-tions Diagnostics, Cergy-Pontoise, France) and venous EDTA-anticoagulated blood specimens by the central laboratory high-performance liquid chromatogra-phy measurement (Tosoh HLC-723