Quantitative Stability of DNA after Extended Storage of Clinical Specimens as Determined by Real-Time PCR
Year: 2002
Type: article
Abstract: Viral DNA stored for extended periods can be amplified by PCR. However, it is unknown whether stored specimens give accurate quantitative results by newer real-time PCR techniques. We therefore compared herpes simplex virus DNA levels in specimens before and after 16 months of storage. The levels of viral DNA remained stable whether the DNA was stored as purified DNA or unextracted DNA in a whole specimen.
Source: Journal of Clinical Microbiology
Language: English
Subfield: Molecular Biology
Domain: Life Sciences
Open Access status: green
OpenAlex ID: W2091966037