Progression of human bone ingrowth into porous-coated implants: Rate of bone ingrowth in humans
Year: 1997
Type: article
Abstract: We report the measured progression of human cancellous bone ingrowth into load-bearing porous-coated titanium implants over 5 time periods (0, 3, 6, 9, and 12 months). There was a statistically significant progression of bone ingrowth into the implants over a 9-month period, but the 9- and 12-month data were not different. Investigators are advised to analyze time "0" implants in order to distinguish mechanical impaction of bone from the biological process of bone ingrowth.
Source: Acta Orthopaedica Scandinavica
Authors: Aaron A Hofmann, Roy D Bloebaum, Kent N Bachus
Language: English
Subfield: Surgery
Field: Medicine
Domain: Health Sciences
Open Access status: gold