Work

Square Wave Cathodic Stripping Voltammetry Adsorptive for Nickel and Cobalt Analysis

Year: 2010
Type: article
Abstract: The adsorptive stripping voltammetric determination of Ni and Co based on adsorption of the Ni/Co and dimethylglioxime (DMG) complex on a hanging mercury drop electrode is studied. The reduction current of the adsorbed DMG complex is measured by square wave cathodic stripping voltammetry method. The effect of various parameters such as ligand concentration, pH of supporting electrolytic, adsorption potential and adsorption time on the current peak of Ni and Co voltammogram were studied. Optimum condition of this method are supporting electrolyte pH 9, DMG concentration 5×10 -4 M, adsorption potential -0.7 V vs Ag/AgCl and adsorption time 180 second. A linier relationship between the current peak and Ni or Co concentration was obtained in the range 5 - 30 ng/mL and the detection limit 0.6 ng/ml for both Ni and Co. The recovery of Ni and Co were 98.11-104.17% using standard biological materials with RSD 2.59 - 10.37%. Based on ";t"; test can be conclude that the result are nearly equal to the standard reference material. Keywords: adsorptive stripping voltammetric, dimethylglioxime complex, nickel, cobalt
Authors: Saryati Saryati, Sumardjo Sumardjo
Language: English
Field: Chemistry
Open Access status: gold
Estimated APC paid: $300
OpenAlex ID: W2101936480
FWCI:
Cites:
Cited by:
Related to: 10