Work

Tolerance within a surgical guide

Year: 2010
Type: article
Abstract: BACKGROUND: Recently, guided surgery has been introduced and several papers verified its accuracy. AIM: While those studies reported on the accuracy of the entire procedure, this experiment wanted to make clinicians aware of the amount of the deviation that can already occur during the drilling procedure alone, due to the tolerance of the drill in the drill key. MATERIAL AND METHODS: Drilling was executed in plexi-glass with a maximal inclination of the drills within the drill keys. RESULTS: A mean deviation in angulation of 4.7 degrees occurred with a mean horizontal deviation at the implant shoulder of 0.8 mm and 1.8 mm at the apex of 13 mm implant. The deviation was further dependent of the implant length, the distance of the sleeve above the bone and the respective guiding system. DISCUSSION: Increasing the height of the drill key will minimize the inaccuracy. CONCLUSIONS: The results from this experiment showed the importance of keeping the drill parallel to the guide in a centric position.