Year: 1982
Type: article
Abstract: A homogenous enzyme immunoassay for the quantitative determination of netilmicin in serum was developed. The procedure utilizes a commercially available assay for gentamicin (EMIT; Syva Co., Palo Alto, Calif.). The method was adapted to a microcentrifugal analyzer, and log-logit regression analysis was performed with a computer. The results of samples assayed by this method correlate well with microbioassay (r2 = 0.985) and radioimmunoassay (r2 = 0.986). This method is not only precise and accurate, but also very rapid and economical and compares favorably to other available methods of netilmicin assay.
DOI: 10.1128/aac.21.3.399
Authors: T Larson, D N Gerding, L R Peterson, J H Eckfeldt
Language: English
Subfield: Pharmacology
Field: Medicine
Domain: Health Sciences
Open Access status: bronze
OpenAlex ID: W2111993597