Work

Fiber optic sensing of relative humidity using a twin low coherence interferometer

Year: 2010
Type: article
Abstract: A fiber-optic implementation of psychrometer is reported. It consists of two Sagnac interferometers with twisted highly birefringent fiber that is used as a temperature sensitive element. One interferometer is used for dry-bulb temperature measurement and the other is for wet-bulb temperature monitoring. The interrogation technique for low coherence interferometric sensor with signal processing in spectral domain is described. The measurement uncertainties for temperature and relative humidity measurements no worse than 0.01°C and 4%, respectively, have been achieved experimentally.
Language: Catalan
Open Access status: diamond
OpenAlex ID: W2114255771
FWCI: 0.42
Cites: 4
Cited by: 2
Related to: 10