Quantitative high-resolution measurement of cerebrovascular physiology with slip-ring CT.
Year: 1996
Type: article
Abstract: PURPOSE: To implement and validate spiral slip-ring CT for use in cerebrovascular studies. METHODS: Continuous data were acquired from an experimental, first-pass, iodine contrast, bolus study by unidirectional X-ray tube rotation, and images were reconstructed at 100-millisecond intervals. Functional maps of cerebral blood volume (CBV) and cerebral blood flow (CBF) were constructed with voxel-by-voxel gamma variate fitting. Reproducibility studies, different injection volumes and sites, and CO2 challenge were applied to verify the technique. RESULTS: Average absolute cortical gray and white matter and basal ganglia results were reproducible within +/- 0.8 ml/100 g for CBV and +/- 20 ml/100 g per minute for CBF, CBV response to changing arterial CO2 tension was significant only in cortical gray matter and basal ganglia; CBF response was significant in gray and white matter, as well as in the basal ganglia. CONCLUSION: Functional CT and constructed functional maps provide an optimal, high-resolution tool with which to visualize cerebrovascular parameters and their changes.
Source: PubMed
Authors: L M Hamberg, G J Hunter, E F Halpern, B Hoop, G S Gazelle, G L Wolf
Institution: Massachusetts General Hospital
Language: English
Subfield: Radiology, Nuclear Medicine and Imaging
Field: Medicine
Domain: Health Sciences
Open Access status: green
OpenAlex ID: W2114700987