An image processing approach to calibration of hydrometers

Year: 2004
Type: article
Abstract: The usual method adopted for multipoint calibration of glass hydrometers is based on the measurement of the buoyancy by hydrostatic weighing when the hydrometer is plunged in a reference liquid up to the scale mark to be calibrated. An image processing approach is proposed by the authors to align the relevant scale mark with the reference liquid surface level. The method uses image analysis with a data processing technique and takes into account the perspective error. For this purpose a CCD camera with a pixel matrix of 604H × 576V and a lens of 16 mm focal length were used. High accuracy in the hydrometer reading was obtained as the resulting reading uncertainty was lower than 0.02 mm, about a fifth of the usual figure with the visual reading made by an operator.
Source: Metrologia