Year: 2012
Type: article
Abstract: In high static field magnetic resonance imaging(MRI) systems, $B_0$ fields of 7 T and 9.4 T, the impressed RF field shows larger inhomogeneity than in clinical MRI systems with B0 fields of 1.5 T and 3.0 T. In multi-channel RF coils, the magnitude and phase of the input to each coil element can be controlled independently to reduce the non-uniformity of the impressed RF $B_1^+$ field. The convex optimization technique has been used to obtain the optimum excitation parameters with iterative solutions for homogeneity in a selected ROI(Region of Interest). To demonstrate the technique, the multichannel transmission line coil was modeled together with a human head phantom at 400 MHz for the 9.4 T MRI system and $B_1^+$ fields are obtained. In this paper, all the optimized $B_1^+$ in each isolated ROIs are combined to achieve significantly improved homogeneity over the entire field of view. The simulation results for 9.4 T MRI systems are discussed in detail.