Work

Dual-mode IVUS catheter for intracranial image-guided hyperthermia: Feasibility study

Year: 2010
Type: article
Abstract: In this study, we investigated the feasibility of modifying 3-Fr IVUS catheters in several designs to potentially achieve minimally-invasive, endovascular access for image-guided ultrasound hyperthermia treatment of tumors in the brain. Using a plane wave approximation, target frequencies of 8.7 and 3.5 MHz were considered optimal for heating at depths (tumor sizes) of 1 and 2.5 cm, respectively. First, a 3.5-Fr IVUS catheter with a 0.7-mm diameter transducer (30 MHz nominal frequency) was driven at 8.6 MHz. Second, for a low-frequency design, a 220-μm-thick, 0.35 x 0.35-mm PZT-4 transducer--driven at width-mode resonance of 3.85 MHz--replaced a 40-MHz element in a 3.5-Fr coronary imaging catheter. Third, a 5 x 0.5-mm PZT-4 transducer was evaluated as the largest aperture geometry possible for a flexible 3-Fr IVUS catheter. Beam plots and on-axis heating profiles were simulated for each aperture, and test transducers were fabricated. The electrical impedance, impulse response, frequency response, maximum intensity, and mechanical index were measured to assess performance. For the 5 x 0.5-mm transducer, this testing also included mechanically scanning and reconstructing an image of a 2.5-cm-diameter cyst phantom as a preliminary measure of imaging potential.
Language: English
Open Access status: green
OpenAlex ID: W2126157357
FWCI: 1.132
Cites: 51
Cited by: 13
Related to: 10