Maximum-likelihood calibration of an x-ray computed tomography system
Year: 2010
Type: conference-paper
Abstract: We present a maximum-likelihood (ML) method for calibrating the geometrical parameters of an x-ray computed tomography (CT) system. This method makes use of the full image data and not a reduced set of data. This algorithm is particularly useful for CT systems that change their geometry during the CT acquisition, such as an adaptive CT scan. Our ML search method uses a contracting-grid algorithm that does not require initial starting values to perform its estimate, thus avoiding problems associated with choosing initialization values.
Authors: J W Moore, R Van Holen, H H Barrett, L R Furenlid
Institutions: University of Arizona, Ghent University
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W2129392987