Practical parallel imaging compressed sensing MRI: Summary of two years of experience in accelerating body MRI of pediatric patients

Year: 2011
Type: conference-paper
Abstract: For the last two years1, we have been experimenting with applying compressed sensing parallel imaging for body imaging of pediatric patients. It is a joint-effort by teams from UC Berkeley, Stanford University and GE Healthcare. This paper aims to summarize our experience so far. We describe our acquisition approach: 3D spoiled-gradient-echo with poisson-disc random undersampling of the phase encodes. Our reconstruction approach: ℓ1-SPIRiT, an iterative autocalibrating parallel imaging reconstruction that enforces both data consistency and joint-sparsity in the wavelet domain. Our implementation: an on-line parallelized implementation of ℓ1-SPIRiT on multi-core CPU and General Purpose Graphics Processors (GPGPU) that achieves sub-minute 3D reconstructions with 8-channels. Clinical results showing higher quality reconstruction and better diagnostic confidence than parallel imaging alone at accelerations on the order of number of coils.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W2130299261
FWCI: 26.97
Cites: 8
Cited by: 172
Related to: 10