Work

Influence of Off-resonance in myocardial T1-mapping using SSFP based MOLLI method

Year: 2013
Type: article
Abstract: BACKGROUND: Myocardial T1-mapping methods such as MOLLI use SSFP readout and are prone to frequency-dependent error in T1-measurement. A significant error in T1 may result at relatively small off-resonance frequencies that are well within the region without banding artifacts. METHODS: The sensitivity of T1-estimates based on the SSFP based MOLLI sequence to errors in center frequency are calculated by means of a Bloch simulation and validated by phantom measurements. Typical off-resonance errors following local cardiac shimming are determined by field mapping at both 1.5 and 3.0T. In vivo examples demonstrate the artifactual appearance of T1-maps in the presence of off-resonance variation. RESULTS: Off-resonance varied 61.8 ± 15.5 Hz (mean ± SD, n = 18) across the heart at 1.5T and 125.0 ± 40.6 Hz (mean ± SD, n = 18) at 3.0T. For T1 = 1000 ms, the variation in T1 due to off-resonance variation was approximately 20 ms at 62 Hz, and > 50 ms at 125 Hz. CONCLUSIONS: Regional variations due to the inability to completely shim the B0-field variation around the heart appear as regional variation in T1, which is artifactual.
Language: English
Field: Medicine
Open Access status: gold
Estimated APC paid: $3,000
OpenAlex ID: W2131786529
FWCI: 8.556
Cites: 24
Cited by: 129
Related to: 10