Assessment of the Performance of the Adaptive Thresholding Algorithm for QRS Detection with the Use of AHA Database

Year: 2007
Type: article
Abstract: Two modifications of an adaptive thresholding algorithm for heart beat detection have already been developed. The threshold combines three parameters: an adaptive slewrate value, a second value which rises when high-frequency noise occurs, and a third one intended to avoid missing of low amplitude beats. The current study assesses the performance of a new modification of the combined adaptive thresholding method for heart beat detection with the use of AHA database. The results are: Mod. 1 Se=99.58%, Sp=99.83 %; Mod. 2 Se=99.73%, Sp=99.83%; Mod. 3 Se=99.78%, Sp=99.85%. The statistical indices are higher than, or comparable to those, cited in the scientific literature.