Work

Design and Calibration of the EPA PM2.5 Well Impactor Ninety-Six (WINS)

Year: 2001
Type: article
Abstract: The EPA well impactor ninety-six (WINS) was designed and calibrated to serve as a particle size separation device for the EPA reference method sampler for particulate matter under 2.5  m aerodynamic diameter. The WINS was designed to operate downstream of a PM 10 inlet at a volumetric flow rate of 16.7 Lpm. For design simplicity and ease of construction, fractionation of the aerosol in the WINS is provided by a single-stage, single-jet, round-hole impactor. Particles > 2.5  m (aerodynamic diameter) have sufficient inertia to be impacted upon a circular 37 mm diameter glass fiber filter immersed in 1 mL of a low volatility oil. The relatively large amount of oil is intended to minimize substrate overloading and subsequent particle bounce experienced by some conventional impactors and to represent an easily field dispensable quantity of defined tolerance. The novel geometry of the impaction reservoir (or well) is designed to capture any reentrained material fromthe impaction surface and to prevent loss of oil should the unit be inadvertently turned over or onto its side. The penetration curve of the final WINS design has a 50% cutpoint diameter equal to 2.48  m and a geometric standard deviation of 1.18. During development, several nozzle designs and well geometries were evaluated to optimize the performance of the WINS. Additionally, two candidate oils (Neovac and Dow Corning 704 diffusion pump oils) and three types of filters (glass fiber filters, drain discs, and polycarbonate membrane filters) were evaluated for use as impaction substrates in the WINS. The performance of the WINS was similar for the two oils in combination with a glass fiber filter and a drain disc; however, a polycarbonate filter demonstrated elevated penetration values. Based on these tests, a Gelman Type A/E glass fiber filter immersed in Dow Corning 704 diffusion pump oil was selected as the best impaction substrate. Further testing showed that the penetration curve was essentially the same when operated with quantities of oil ranging from 0.75 to 3 mL.
Language: English
Open Access status: bronze
OpenAlex ID: W2147698518
FWCI: 2.125
Cites: 16
Cited by: 76
Related to: 10