Hydrogen maser ensemble performance and characterization of frequency standards
Year: 2003
Type: conference-paper
Abstract: The performance of a post-processed clock ensemble dominated by active, cavity-tuned, hydrogen masers has been evaluated. The ensemble has an average frequency drift of less than /spl plusmn/3/spl times/10/sup -15/ per year, and an Allan deviation less than 1/spl times/10/sup -15/ for /spl tau/ between 4 hours and 100 days. This ensemble has been used to help characterize the frequency stability of a number of commercially available, high performance frequency standards.
Author: T.E. Parker
Institution: National Institute of Standards and Technology
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W2155363583