Year: 2007
Type: article
Abstract: A reflection-mode photoacoustic (PA) imaging system was designed and built to image deep structures in biological tissues. We chose near-infrared laser pulses of 804-nm wavelength for PA excitation to achieve deep penetration. To minimize unwanted surface signals, we adopted dark-field ring-shaped illumination. This imaging system employing a 5-MHz spherically focused ultrasonic transducer provides penetration up to 38 mm in chicken breast tissue. At the 19-mm depth, the axial resolution is 144 microm and the transverse resolution is 560 microm. Internal organs of small animals were imaged clearly.
Source: Journal of Biomedical Optics
DOI: 10.1117/1.2818045
Authors: Kwanghyun Song, Lihong V. Wang
Institution: Washington University in St. Louis
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: bronze
OpenAlex ID: W2156906835