Measuring Markers of Liver Function Using a Micropatterned Paper Device Designed for Blood from a Fingerstick
Year: 2012
Type: article
Abstract: This paper describes a paper-based microfluidic device that measures two enzymatic markers of liver function (alkaline phosphatase, ALP, and aspartate aminotransferase, AST) and total serum protein. A device consists of four components: (i) a top plastic sheet, (ii) a filter membrane, (iii) a patterned paper chip containing the reagents necessary for analysis, and (iv) a bottom plastic sheet. The device performs both the sample preparation (separating blood plasma from erythrocytes) and the assays; it also enables both qualitative and quantitative analysis of data. The data obtained from the paper-microfluidic devices show standard deviations in calibration runs and "spiked" standards that are acceptable for routine clinical use. This device illustrates a type of test useable for a range of assays in resource-poor settings.
Source: Analytical Chemistry
DOI: 10.1021/ac203434x
Institutions: Harvard University, Diagnostics For All
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: green
OpenAlex ID: W2161314076