Work

Insulation Verification Using Low Voltage and High Current Sensitivity

Year: 2012
Type: article
Abstract: The classic approach for determining insulating resistance by using commercial testers is not reliable for testing different types of insulation, especially for cables with one conductor and nanoelectronic oxide-based devices (nanotransistors).In this paper an alternative low-voltage method for measuring insulation resistance based on strongly increased current sensitivity is proposed.The proposed method of measuring insulation resistance at low voltages is realized by using a high sensitivity Keithley SMU 2612 unit.The verification of the proposed method is made by measuring insulation resistance of medical Corkscrew Electrode and gate oxide resistance for nano-scaled MOSFET.Finally, results of measuring insulation resistance obtained by commercial testers and proposed method in the verification process are compared.
Language: English
Open Access status: green
OpenAlex ID: W2161721983
FWCI:
Cites: 4
Cited by:
Related to: 10