Year: 2012
Type: article
Abstract: China Spallation Neutron Source (CSNS), one project of the 12 th Five-Year-Plan scheme of China, is under construction in Guangdong province. Three neutron spectrometers will be installed during the first phase of the project, and two-dimensional position sensitive thermal neutron detectors are required. Before the construction of the neutron detectors, a prototype of a two-dimensional 200 mm × 200 mm Multi-wire Proportional Chamber (MWPC) with Ar/CO 2 (90/10) flowing gas has been constructed. In 2009, the prototype was tested with the 55 Fe X-ray using part of the electronics, and performed well. Following the test in 2009, the neutron detector was constructed with the complete electronics and filled with the 6 atm. 3 He+2.5 atm. C 3 H 8 gas mixture in 2010. The neutron detector has been primarily tested with an Am-Be source. In this paper, new developments of the neutron detector including the design of the high pressure chamber, the optimization of the gas purifying system and the gas filling process will be reported. The results and discussion are also presented in this paper.
Source: Chinese Physics C
Authors: Li-Chao Tian, Bin Tang, Xiao-Hu Wang, Rong-Guang Liu, Jian Zhang, Yuan-Bo Chen, Zhi-Jia Sun, Hong Xu, Gui-An Yang, Qiang Zhang
Language: English
Subfield: Nuclear and High Energy Physics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: bronze
Award: 二维位置灵敏涂硼GEM中子探测器
OpenAlex ID: W2220892409