Work

Three‐dimensional MR spectroscopic imaging using adiabatic spin echo and hypergeometric dual‐band suppression for metabolic mapping over the entire brain

Year: 2016
Type: article
Abstract: PURPOSE: Large lipid and water signals in MR spectroscopic imaging (MRSI) complicate brain metabolite quantification. In this study, we combined adiabatic hypergeometric dual-band (HGDB) lipid and water suppression with gradient offset independent adiabatic (GOIA) spin echo to improve three-dimensional (3D) MRSI of the entire brain. METHODS: 3D MRSI was acquired at 3T with a 32-channel coil. HGDB pulses were used before excitation and during echo time. A brain slab was selected with GOIA-W(16,4) pulses, weighted phase encoded stack of spirals, and real-time motion/shim correction. HGDB alone or in combination with OVS and MEGA (MEscher-GArwood) was compared with OVS only and no suppression. RESULTS: The combined HGDB pulses suppressed lipids to 2%-3% of their full unsuppressed signal. The HGDB lipid suppression was on average 5 times better than OVS suppression. HGDB+MEGA provided 30% more suppression compared with a previously described HGDB+OVS scheme. The number of voxels with good metabolic fits was significantly larger in the HGDB data (91%-94%) compared with the OVS data (59%-80%). CONCLUSION: inhomogeneity. Magn Reson Med 77:490-497, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
Language: English
Field: Medicine
Open Access status: green
OpenAlex ID: W2263002615
FWCI: 2.311
Cites: 40
Cited by: 19
Related to: 10