Work

The scanning transmission microscope at the NSLS (National Synchrotron Light Source)

Year: 1989
Type: article
Abstract: The scanning transmission soft x-ray microscope (STXM), that has been under development at the National Synchrotron Light Source has been substantially upgraded for operation with the X1 undulator. The principal new features are: optical prefocusing, using a visible light interferometer, a dedicated VAXstation 3200 with a more user friendly and flexible software system for image acquisition and analysis, a flow cell that makes it possible not only to keep the specimen wet during exposure, but to change the fluid around the specimen as well, and a more compact proportional counter that is capable of counting rates of several MHz. In conjunction with new zone plates of better resolution and higher efficiency, the microscope is ready for a period of extended use in biological imaging. 9 refs., 6 figs.
Language: English
Subfield: Radiation
Open Access status: green
OpenAlex ID: W2284774572
FWCI:
Cites: 5
Cited by:
Related to: 10