Safety and Performance Considerations for Interactions Between Electromagnetic Fields and Biological Tissue: Applications to High Field Human Magnetic Resonance Imaging and Tissue-implanted Devices
Year: 2010
Type: article
Abstract: The principal advantage of magnetic resonance imaging (MRI) at high field is the increase in signal to noise ratio (SNR), however, high field imaging also leads to an increased Larmor (operating) frequency, thus the wavelength in tissue can become comparable to the size of the load and/or the coil. The performance of the radiofrequency (RF) coil, as a result, becomes increasingly dependent on its electromagnetic interactions with the load, human body or head.