A computer-supported oxystat system maintaining steady-state O2 partial pressures and simultaneously monitoring O2 uptake in biological systems
Year: 1986
Type: article
Abstract: A feedback-controlled oxystat system is described maintaining steady-state O2 partial pressures (pO2) between 0.01 mmHg (14 nM-O2) and 150 mmHg (210 microM-O2) and simultaneously monitoring O2 uptake at rates between 0.1 and 120 microM-O2 X min-1 in suspensions of cells, in subcellular fractions and in solutions of enzymes. At pO2 values between 0.2 and 150 mmHg (0.28 and 210 microM-O2) a polarographic O2 sensor was used, and below a pO2 of 0.2 mmHg (0.28 microM-O2) the O2-dependent luminescence of the photobacterium Vibrio fischeri was utilized to monitor the actual pO2. At a selected pO2, O2 supply is maintained by injecting appropriate amounts of O2-saturated aqueous medium into the reaction chamber by using a motor-driven burette. The oxystat system is under control of a computer that reads the O2 sensors, interacts with the motor-driven burette, calculates the O2 uptake from the amounts of O2-saturated medium added, collects data from further measuring devices and provides the documentation of the results during incubation.
Source: Biochemical Journal
DOI: 10.1042/bj2360765
Authors: T Noll, H de Groot, P Wissemann
Institution: Heinrich Heine University Düsseldorf
Language: English
Subfield: Analytical Chemistry
Field: Chemistry
Domain: Physical Sciences
Open Access status: bronze