Abstract: Laboratory experiments were performed to adapt a catalytic colorimetric \nmethod to continuous flow analysis for the shipboard determination of iron(III) in \nseawater. System parameters such as reagent concentrations, reaction temperature, \nflow rate, reaction coil length, pH and buffer grade were optimized to increase \nsensitivity and decrease background absorbance. The method is based on the catalytic \neffect of iron(III) on the oxidation of N, N-dimethyl-p-phenylenediamine by hydrogen \nperoxide, and has an optimal detection limit of 0.2 nM. Preliminary shipboard testing \nof the system was conducted to determine if this method would be tactically useful \nwhen applied to a near real time chemical mapping strategy. A mesoscale fertilization \nexperiment, IronEx-1, was performed in the equatorial Pacific in October, 1993. Data \nwere collected over a 24 hr. period following iron enrichment. Iron concentrations \nmeasured by this method compared well with those determined by an independent, \nsemicontinuous chemiluminescence method using samples collected at the same time.