RF induced heating of pacemaker/ICD lead-tips during MRI Scans at 1.5T and 3T: evaluation in cadavers
Year: 2016
Type: article
Abstract: Studies about the safety of MRI exams for patients with pacemakers/ICDs at 1.5T have been reported [Nazarian S., et al. Annals of internal medicine 155.7 (2011): 415-424.]. Of most concern is possible heating of the lead-tips in contact with the myocardium. Little is known about the relative safety of 3T MRI exams for these patients and in vivo lead-tip heating data is difficult, if not impossible, to obtain. Our objective was to measure lead-tip heating directly in human cadavers with pacemakers/ICDs at both 3.0T and 1.5T. Cadavers (N = 5, 3 male, Table 1 , part 1) with existing pacemakers had fiberoptic temperature probes implanted adjacent to right atrial (RA), right ventricular (RV) and/or abandoned lead-tips under x-ray guidance. Whole-body CT was used to estimate lead-tip to probe-tip distances. Cadavers were exposed to 15-minutes of 4 W/kg whole body SAR at both 1.5T and 3T (Siemens Avanto and Prisma) for five isocenter positions: 6 cm superior to the chin (LM 1 ) and in four 15 cm increments inferior to LM 1 (LM 2 to LM 5 ) in order to evaluate lead-tip heating as a function of the device's position relative to isocenter (i.e. different MRI exams). Maximum temperature increases at the lead-tip were reported as △ T Max (temperature difference between the baseline before the MRI sequence and peak heating after 15-minutes). All temperature probes were ≤10 mm from the lead-tip (Table 1 , part 2). Figure 1 shows △ T Max for 1.5T and 3T for each probe. Maximum heating was observed at the LM 2 and LM 3 isocenter positions for which the whole implant was inside the body transmit coil. △ T Max was >2C for 7 of 25 cases at 1.5T and for 12 of 25 cases at 3T, but never exceeded 4C. △ T Max for each isocenter position . Cadavers 1 and 3 have single chamber pacemakers (RV leads); cadavers 2, 4, and 5 have dual chamber pacemakers (RA and RV lead). Cadaver 5 also had 10 abandoned leads. For single chamber pacemakers a temperature probe was placed at the lead-tip and the remaining temperature probes were placed in remote tissue for reference. For dual chamber pacemakers, the temperature probes were placed at the RA and RV lead-tips and two probes were placed in remote tissue for reference. For the cadaver with abandoned leads two temperature probes were placed close to the lead-tips connected to the pacemaker and one probe was placed close to lead-tip of one abandoned lead (ALT). Note that in cadaver 2 the RV temperature probe was not close enough to the RV lead-tip (accidentally partially withdrawn) to obtain temperature data. These data do not indicate a substantial difference between lead-tip heating at 1.5T and 3T, nor do they indicate △ T Max >4C despite the lack of cooling due to tissue perfusion. Continued evaluation is warranted and on-going.
Institution: University of California, Los Angeles
Language: English
Subfield: Cardiology and Cardiovascular Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: gold
Estimated APC paid: $3,000
OpenAlex ID: W2320848213