Work

Rotational distortion correction in endoscopic optical coherence tomography based on speckle decorrelation

Year: 2015
Type: article
Abstract: We present a new technique for the correction of nonuniform rotation distortion in catheter-based optical coherence tomography (OCT), based on the statistics of speckle between A-lines using intensity-based dynamic light scattering. This technique does not rely on tissue features and can be performed on single frames of data, thereby enabling real-time image correction. We demonstrate its suitability in a gastrointestinal (GI) balloon-catheter OCT system, determining the actual rotational speed with high temporal resolution, and present corrected cross-sectional and en face views showing significant enhancement of image quality.
Language: English
Open Access status: green
OpenAlex ID: W2325496248
FWCI: 3.145
Cites: 16
Cited by: 66
Related to: 10