Work

Methodological Considerations in the Determination of Plasma Catecholamines by High-Performance Liquid Chromatography with Electrochemical Detection

Year: 1985
Type: article
Abstract: We report on a systematic investigation into some frequently encountered problems in the estimation of catecholamines in plasma and biological fluids using high-performance liquid chromatography with electrochemical detection. The kinetics of adsorption and desorption of catecholamines from plasma on to alumina has been studied using laboratory-prepared and some commercial aluminas with various acids. Chromatographic conditions yielding optimal resolution and sensitivity have been characterised. Some possible interfering peaks have been identified. The stability of catecholamines in plasma has been studied.
Language: English
Field: Chemistry
Open Access status: bronze
OpenAlex ID: W2327298537
FWCI: 7.506
Cites: 35
Cited by: 164
Related to: 10