Pulse oximetry monitoring of sedated pediatric dental patients.
Year: 1985
Type: article
Abstract: Hypoxemia is recognized as a major complication of sedating pediatric dental patients. Traditional methods of patient monitoring show changes only with moderate to severe hypoxemia. This study compared pulse oximetry, a new monitoring technology, to traditional techniques in their ability to detect hypoxemia in sedated children. The results demonstrated that pulse oximetry is a more sensitive monitor of mild to moderate hypoxemia than measurements of heart rate, blood pressure, respiratory rate, or visual observation for cyanosis in sedated pediatric dental patients.
Source: PubMed Central
Authors: W A Mueller, J N Drummond, T A Pribisco, R F Kaplan
Language: English
Subfield: Anesthesiology and Pain Medicine
Field: Medicine
Domain: Health Sciences
Open Access status: green