Year: 2015
Type: article
Abstract: The massive outbreak of highly lethal Ebola hemorrhagic fever in West Africa illustrates the urgent need for diagnostic instruments that can identify and quantify infections rapidly, accurately, and with low complexity. Here, we report on-chip sample preparation, amplification-free detection and quantification of Ebola virus on clinical samples using hybrid optofluidic integration. Sample preparation and target preconcentration are implemented on a PDMS-based microfluidic chip (automaton), followed by single nucleic acid fluorescence detection in liquid-core optical waveguides on a silicon chip in under ten minutes. We demonstrate excellent specificity, a limit of detection of 0.2 pfu/mL and a dynamic range of thirteen orders of magnitude, far outperforming other amplification-free methods. This chip-scale approach and reduced complexity compared to gold standard RT-PCR methods is ideal for portable instruments that can provide immediate diagnosis and continued monitoring of infectious diseases at the point-of-care.
Source: Scientific Reports
DOI: 10.1038/srep14494
Language: English
Subfield: Biomedical Engineering
Field: Engineering
Domain: Physical Sciences
Open Access status: gold
Estimated APC paid: $2,690
Awards: G1306769321, Collaborative Research: Ultrasensitive Cancer Biomarker Detection on Biophotonic Chips, G2876299153, Hybrid integrated molecular analysis (HIMAS) for point-of-care diagnostics, Hybrid integrated molecular analysis (HIMAS) for point-of-care diagnostics, Graduate Research Fellowship Program, G8319073317, Collaborative Research: Ultrasensitive Cancer Biomarker Detection on Biophotonic Chips
OpenAlex ID: W2415054003