Development of a force displacement measurement device for the determination of spring constants
Year: 2014
Type: article
Abstract: This paper discusses a measurement device for the determination of force displacement curves or spring constants respectively.Especially the calibration of spring constants of atomic force microscopes (short: AFM) cantilevers is an important field of investigation and concentration of this work.The spring constant can be measured with two measurement modes.One mode uses separate force and displacement measurement devices which correspond to the state of the art.The second measurement mode is more sophisticated using one sensor which measures force as well as the displacement simultaneously; a separate nanostage is not need.The measured sample is an AFM type cantilever, its determined spring constant is in both measurement modes approximately 50 Nm -1 with a very good repeatability of 0.02 %.
Language: English
Subfield: Atomic and Molecular Physics, and Optics
Field: Physics and Astronomy
Domain: Physical Sciences
Open Access status: green